The station

Product Name: Anesthesia
Product Type: S6100
Product Introduction:

 S6100 Anesthesia system 

Intended Application

S6100 anesthesia system used for medical service for patients with inhalation anesthesia, controlled breathing, or helps to breathe as well as monitoring and display of the patient's ventilation parameters.

Adaptation disease

? Various patients who need anesthetization.

Specifications

? Adopts the electric control gas driven mode.

? With a variety of alarm, presentation and protection function

? With backup power supply. When the power is automatically switched to backup power

? Use 10.4 "color LCD display shows ventilation parameters, alarm information, and the graphical display.

? For adults and children

? With volume control, pressure limit and work in a variety of ways

? Using temperature and flow compensation and zero self-locking anesthetics vaporizer;

? O2, N2O flowmeter has the linkage,ensure O2 concentration≥25%;

? With O2 drive insufficient pressure alarm and N2O truncation protection

? Equipped with exhaust gas transmission and collection system

Working environment

? Air source                                            O2、N2O、AIR

? Air pressure 280 kPa ~ 600 kPa

? Power supply                                   100-240V~

? Frequency                                          50 /60HZ

? Input power                                          80 VA

 

Ventilation mode

? IPPV

? A/C

? PCV       

? SIMV

? SIGH

? MANU

Main technical parameters

Adjustable parameters

? Flow gauge

O2                             0.1~10 L/min

N2O                              0.1~10 L/min

AIR                       0.1~10 L/min

? Oxygen flush 35 L/min~ 75 L/min

? Tidal volume(Vt) 0 mL ~ 1500 mL

?Including:                            

?For adult 250 mL ~ 1500 mL

?For children 0 mL~ 300 mL

? MV                             

  For adult≥18 L/min

  For children≥10 L/min

? Frequence(Freq)               1 bpm ~ 100 bpm

(SIMV mode:1 bpm~ 40 bpm Except SIMV mode :4 bpm~ 40 bpm for adult ;20bpm~ 100 bpm for children )

? I/E 4:1~1:8

? PEEP                        0 cmH2O ~ 30 cmH2O

? Stress triggering sensitivity(PTr)-20 cmH2O ~ 0 cmH2O(Based on PEEP)

? Flow trigger sensitivity (FTr)       closed, 0.5 L/min ~ 30 L/min

? Pressure control(Pc)               5 cmH2O ~ 60 cmH2O

? Breathless time (aspiratory breathless, breath at the end of the platform)(TIP)0 ~ 6s(≤50 % Aspiratory time)

? SIGH 1/100 ~ 5/100

? Apnea Ventilation OFF,5 s ~ 60 s

? Apnea Ventilation        20 cmH2O ~ 100 cmH2O

Monitored Parameters

? Frequency(Freq)                      0 /min ~ 100 /min

? Tidal volume(Vt) 0 mL ~ 2000 mL

? MV                               0 L/min ~ 100 L/min

? airway pressure 0 cmH2O ~ 100 cmH2O

? Patients with dynamic lung compliance 1mL/cmH2O ~ 1000 mL/cmH2O

? Concentration of inspired oxygen 15 % ~ 100 %

Graphical display:

? P-T

? F-T

? P-V loop

Alarm and protection

? The AC power outage alarm

? Internal battery backup low voltage alarm

? No tidal volume

? High (low) ventilation alarm

? High (low)Airway pressure alarm

? Pressure limit

? Continuous pressure alarm

? High (low)Oxygen concentration alarm

? Suffocation warning

? The maximum limited pressure <12.5 kPa

? With O2 drive insufficient pressure alarm and N2O truncation protection

? O2, N2O flowmeter has the linkage to ensure O2 concentration≥25%。When O2 closed ,N2O closes at the same time.

 Note: the standard one, and optional two at most


Previous:X-ray Machine
Next:S9600 BiPAP CPAP Ventilator info

Related Products