?? Work mode: CPAP, S ,S/T ,T;
?? Air leak compensation: Pressure compensation dynamic monitoring/ Flow compensation dynamic
monitoring;
?? Large color LCD display: It is very convenient to see the setting parameters and statistical data,
meanwhile can display waveform of breathing air and pressure in real time;
?? Multi-parameter Alarm: Mask off alarm, suffocation alarm, high respiratory rate alarm, low respiratory
rate alarm;
Working condition
?? External Power: DC 24 V (-15% ~ +25%), 2.5 A;
?? Power adapter: AC input 100 V -240 V ~, frequency 50HZ / 60 HZ
?? DC output 24 V, 2.5 A.
?? Ambient temperature: 5 ¡æ ~ 35 ¡æ
?? Relative humidity: 15% to 95%
?? Atmospheric pressure 830 hap ~ 1060 HPA
?? Storage Ambient temperature: -20 ¡æ ~ +60;
?? Relative humidity of 15% to 95%
?? Atmospheric pressure 500 HPA ~ 1060 HPA
Ventilation Mode
T, S, S/T, CPAP
Main technical parameters
Ventilation Parameter
?? Aspiratory pressure (IPAP): 4 cmH2O ~ 25 cmH2O
?? Expiratory pressure (EPAP): 4 cmH2O ~ 20 cmH2O
?? Continuous positive airway pressure (CPAP): 4 cmH2O ~ 20 cmH2O
?? Pressure delay rise time : 0 min ~ 60 min
?? Pressure rise time: automatic adjustment, 0.1 s ~ 0.6 s
?? Respiratory rate: 2 ~ 40 BPM
?? Aspiratory time: 0.0 s ~ 3.0 s
?? Aspiratory sensitivity: automatic adjustment, 1, 2, 3;
?? Expiratory sensitivity: automatic adjustment, 1, 2, 3;
?? comfort: Off, 1, 2, 3;
?? Humidifier: 0-5
Contraindications - Absolute contraindications
?? Heart or respiratory arrest;
?? Weak spontaneous breathing, coma;
?? Upper respiratory tract obstruction;
?? Can not clear the oropharynx and upper respiratory tract secretions, respiratory protection ability low;
?? Neck and facial trauma, burns and deformity;
?? Multiple system organ failure (hemodynamic instability, unstable arrhythmia; gastrointestinal
perforation / bleeding, severe brain disease)
Contraindications - Relative contraindications
?? Pneumothorax without drainage;
?? Recent surgery of face, neck, mouth, pharynx, esophagus and stomach;
?? Obvious noncooperation or extreme standstill;
?? Severe hypoxemia (PaO2 <45mmHg) , severe acidosis (pH ¡Ü 7.20);
?? Severe infection;
?? Much respiratory tract secretions or difficult expectoration.
Alarm and protection
?? Detached alarm
?? Suffocation alarm
?? Minute volume alarm
?? Respiratory rate alarm
?? High respiratory tract pressure alarm
?? Power off alarm